1.Large-screen LCD display with controllable white backlight.
2.User-friendly design: featuring a bracket-type structure for convenient use by customers, and an anti-slip glove-friendly design for easy operation.
3.The dissolved oxygen measurement mode is equipped with functions of automatic temperature compensation, salinity compensation and air pressure compensation.
4.The new-structured polarographic dissolved oxygen electrode can measure temperature, salinity and dissolved oxygen simultaneously. Equipped with a dedicated dissolved oxygen electrode calibration sleeve, the electrode polarization only takes 3-5 minutes.
86-175-5409-3886
cedric@cedriccn.com
Specification:
Model | CP610 | |
DO | Measuring range | (0 - 20,00) mg/L(ppm) (0-200.0)% |
Resolution | 0.1/0.01 mg/L(ppm) 1/0.1% | |
Accuracy | Electrometer:±0.10 mg/L, Complete set:±0.40 mg/L | |
Response time | ≤30 s (25℃,90% response) | |
Temperature compensation range | (0 - 99)℃ (Automatic) | |
Salinity compensation range | (0 - 45) ppt | |
Air pressure Compensation range | (0 - 200) kPa | |
Electrode type | Polarographic | |
Other Parameters | Data storage | 150 sets |
Power supply | 2 AA batteries (1.5 V×2) | |
Automatic calibration | Water-saturated air; air-saturated water | |
Dimensions & Weight | Meter:(90×160×30) mm,200 g; Carrying case:(255×210×50) mm,790 g | |
Stored content | Serial number, measurement value, unit of measurement and temperature value | |
Quality and safety certifications | ISO9001:2000, CE and CMC | |
Working Conditions | Temperature | 5 - 35 ℃ (Class 0.01) |
Humidity | ≤85% | |
IP grade | IP57 | |
Fil the form and send us your message.Get in touch with PAULAIR and find a specialized consultant for you !