Tel:86-175-5409-3886 E-mail:cedric@cedriccn.com
UV-VIS UV756
UV-VIS UV756

Instrument Features:1. Adopting 1200/mm high-performance gratings. The new light source control system enables faster switching of instrument light sources.2. The precise 2nm bandwidth makes the test data more accurate. Using imported long-life light sources to reduce the consumption of consumables for users.3. The unique wavelength control system ensures higher wavelength accuracy. Improved optical system for more accurate testing.4. Using imported photoelectric converters to increase the

86-175-5409-3886
cedric@cedriccn.com
Description

Instrument Features:

1. Adopting 1200/mm high-performance gratings. The new light source control system enables faster switching of instrument light sources.

2. The precise 2nm bandwidth makes the test data more accurate. Using imported long-life light sources to reduce the consumption of consumables for users.

3. The unique wavelength control system ensures higher wavelength accuracy. Improved optical system for more accurate testing.

4. Using imported photoelectric converters to increase the sensitivity of the instrument.

5. A new type of microcomputer data processing system. Make the use of the instrument more convenient and stable.

6. The instrument adopts a 126 × 64 bit dot matrix LCD display, which can directly display standard curves and test data. The host can store test data and can be equipped with a printer.

7. The instrument can be optionally equipped with a disc rotating automatic eight link sample holder, which has good reliability and a centered light spot (compared to a double row automatic eight link holder), resulting in high metering accuracy of the instrument.

Automatically save the latest wavelength based on the set wavelength.

Compared to ordinary similar products, it has increased the stability of measurement.

The measurable optical path length is 100mm.

Establish quantitative analysis methods based on standard patterns or input standard patterns.

Can save 200 calibration curves.

You can choose to install software devices that are connected to the computer.


Technical Parameter

Instrument model: UV756 

Wavelength range: 190-1100nm

Wavelength accuracy: ± 0.5nm

Wavelength repeatability: 0.2nm

Spectral bandwidth: 2nm

Transmittance accuracy: ± 0.5% T

Transmission density repeatability: 0.2% T

Baseline straightness: 0.001A/h

Luminous range: -3-3A, 0-200% T, 0-9999C

Stray light: 0.05% T @ 220nm/360nm

Stability: ± 0.001A/h @ 500nm

Display mode: 128 × 64 bit dot matrix LCD display

RS232 communication: USB interface

Printer: optional

Analysis software: Supports

Automatic eight color comparison pool: optional

Optical system: Dual beam proportional detection

Noise level ± 0.001A/ 2min@500nm